Custom CIC
The HF3 wax protection system
To remove a used wax filter:
1. Insert the threaded end of the HF3 tool into the
wax filter on the hearing instrument and then twist
the tool approximately one and a half turns.
2. Pull the HF3 tool out and away from the hearing
instrument. The wax filter should pull out with the
tool. If not, repeat Step #1, with an additional twist
of the tool.
To dispose of a used wax filter:
3.
With the used wax filter attached to the HF3 tool, in-
sert the tool into the center slot of the filter package.
4. Slide the tool to the side. The used wax filter will
remain in the slot when you lift the HF3 tool
straight out.
To install a new wax filter:
5. Press down on a new
wax filter in the pack-
age with the plastic
end of the HF3 tool.
To remove an old wax guard:
1. Insert the removal side of the wax guard tool into
the used wax guard so that the shaft of the tool is
touching the rim of the wax guard.
2. Slowly pull the wax guard tool straight out. Do not
twist the wax guard from the bushing.
To insert a new wax guard:
3. Gently press the replacement side of the wax
guard tool straight into the hole of the sound outlet
until the outer ring lies flush with the outside of your
hearing instrument.
4. Pull the tool straight outthe new wax guard will
remain in place.